Section VII Quality Control

 

The Veterinary Immunofluorescence Analyser has a built-in quality control function that aims to monitor the internal system performance of the meter.

The quality control should be run upon initial use and each day of patient testing.

To run the Quality Control:

  1. Navigate to the main interface and click “QC”. The test device tray will protrude. If the tray does not protrude automatically, click “out”.
  2. Gently insert the Quality Control Card into the test device tray, then push the tray inside the meter until resistance is felt.
  3. Click “Start”.
  4. Wait until the blue progress bar runs to the end. The result of the Quality Control will be displayed.
    • If the result of the quality control is not as expected, contact the distributor to report the issue.
  5. Click “Reset the slot” to retract the test device tray.
  6. Click “Print” to obtain a printed copy of the quality control result if needed.

Section IX Quality Control

The Veterinary Immunofluorescence Analyser has a built-in quality control function that aims to monitor the internal system performance of the instrument.

The quality control should be performed using a quality card to verify initial use and for daily patient testing.

To run the Quality Control:

  1. Navigate to the system interface and click “QC”. The test device tray will protrude. If the test device tray does not protrude automatically, click “Slot Out”.
  2. Gently insert the Quality Control Card into the test device tray, then push the tray inside the instrument until resistance is felt.
  3. Click “QC”.
  4. Wait for the countdown to complete. The QC test is finished, and the result of the Quality Control will be displayed.
    • If the result of the quality control is not as expected, contact the distributor to report the issue.
  5. Click “Slot Reset” to retract the test device tray.
  6. Click “Print” to obtain a printed copy of the quality control result if needed.